logo
Overview | SPECIFICATION

Body Temperature Solution (BTS) - Specification


Body Temperature Scanner - Main Module
Image SensorVanadium Oxide Uncooled Focal Plane Arrays
Max. Resolution160 × 120 (the resolution of output image is 320 × 240)
Pixel Interval17μm
Response Waveband8 μm to 14 μm
NETD (Noise Equivalent Temperature Difference)≤ 40 mk (@25 °C, F# = 1.1)
Lens (Focal Length)6.2 mm
IFOV2.74 mrad
Field of View25° × 18.7° (H × V)
Min. Focusing Distance0.6 m
ApertureF1.1
Temperature Measurement3 temperature measurement rule types, 21 rules in total (10 points, 10 areas, and 1 line).
Temperature Range30°C to +45 °C
Temperature Accuracy±0.5°C
Power Consumption

12 VDC ± 20%: 0.65 A, max. 7.5 W
PoE (802.3af, class 3): 42.5 V to 57 V, 0.14 A to 0.22 A, max. 8 W

Temperature Calibrator
Radiation source temperature5°C ~ 50°C, adjustable
Temperature Resolution0.1°C
Temperature Accuracy+-0.1°C
Temperature Stability+-0.1°C/h
Effective Emissivity0.97+-0.02
Effective Radiation Surface70mm x 70mm
Power Supply220V, AC50Hz
Working Temperature0°C ~ 30°C